Abstract
Women in low- and middle-income countries have limited access to long-acting contraceptives. Access to long-acting contraceptives, such as subcutaneous contraceptive implants, could be increased by task-shifting administration of the implants from skilled to minimally-trained healthcare providers. The objective of this study was to investigate the usability of a task-shifting device for administering subcutaneous contraceptive implants. Healthcare providers (n = 128) from multiple health centers in Ethiopia were trained to administer implants with the existing method and a method using the device on an arm simulator. Participants were observed while inserting implants into the arm simulator, and procedural error rates were calculated. Observations were analyzed using an iterative inductive coding methodology. For the device-assisted method, minimally-trained healthcare providers had a greater procedural error rate than other professions (p = 0.002). For the traditional method, physicians had a greater procedural error rate than nurses and midwives (p = 0.03). Several procedural errors were identified such as participants inserting and removing the trocar and plunger completely and inserting and removing the trocar too far or not enough. These findings reinforce the importance of performing formative usability testing during the early phases of a medical device design process, considering users’ mental models, and avoiding assumptions about healthcare providers’ abilities.